This protocol is to extract RNA and fractionate from 20% 13C-labelled cultures from Siders Pond into 22 fractions.
Ribogreen Quantification
Materials:
- Diluted RNA standard (keep in small aliquots and date – use within 1 month)
- Make 100 ng/ml: Stock is 100g/ml in fridge, dilute 1l:999l in 1X TE
- Make 10 ug/ml: 1:10 dilution of stock, dilute 1l:9l in 1X TE (optional)
- 1X TE diluted from 20X TE provided in kit
- To make 1ml, 50 μl of 20X TE + 950l water
- To make 2ml, 100 μl of 20X TE + 1900l water
- Multiply the number of wells you will be using (including the standard curve by 50, to calculate the number of μL needed of 1X TE and Ribogreen (step 3)
- Black 96-well plate (Costar)
- Pull out Ribogreen to thaw in the dark and make up the Ribogreen solution
Prior to Start:
At least 1 hour before you are ready to read, turn on the computer and open the software for the plate reader (Software: SoftMaxPro). THEN turn on the plate reader. The software will not find the plate reader unless you follow in this order; click on corner icon (SpectraMax), select instrument to connect. During warm-up of plate reader, follow the protocol listed below.
Plan your plate. You will need one well for each of 7 standards (add the number of l equal to the ng/mL; 50 ng/mL = 50 L) and one well for each sample. You can reduce or add standards if you think your nucleic acid falls outside of the standard range. Do not use the edge rows/columns (plate reader may not be accurate here at all edge wells). Also, run duplicates of unfractionated RNA.
Make 1:2000 Ribogreen mixture from aliquot of Ribogreen reagent and 1X TE. E.g. If you need a total of 2 mL of 1:2000 Ribogreen, mix 1 L Ribogreen with 1999 L 1X TE. Protect from light.
Set up plate:
- Add 1X TE to wells in plate. For all samples, add 48 μL 1X TE. For standards, add 50 μL – x 1X TE (for example, for the 5 ng/mL well, add 45 μL). Use table below if needed.
- Use the 1:10 RNA diluted stock (10 ug/ml) for the *2 highest standards in the curve, and then use the 1:1000 (100 ng/ml) for the rest.
| B2 |
49 |
1* |
| C2 |
0 |
50 |
| D2 |
25 |
25 |
| E2 |
40 |
10 |
| F2 |
45 |
5 |
| G2 |
49 |
1 |
| B3 |
50 |
0 |
*use 1:10 – 10 g/ml RNA concentration (ng/mL) Fill rest with nucleic acid 49 1 Determine by standard curve
- Add standard or template according to the table above. Consider adding additional positive and negative controls (e.g., additional standards to run as unknown samples or additional blanks).
- Finally add 50 μL of 1:2000 Ribogreen to each well, pipetting up and down three times to mix. At this step, you’re completing a 1:100 dilution of your RNA
- Adjust settings and fill template editor information:
Settings:
Read Mode: FL
Wavelength: 480-520 nm
Read Type: Well Scan (will take the average of 5 readings in each well)
Plate type: 96 well standard opaque
Read Area: highlight and assign standards, blanks, and sample wells (avoid using outer wells)
Template Editor:
Indicate wells containing standards and enter their concentration
Standard curve will require you to input the concentration but it is in mg/ml, this is OK. Fill in the above values from your standard curve
Also denote the plate blank
Indicate wells containing samples – click assign each time you highlight a well.
Press Read
- It will only take a minute or 2 to read your plate.
- Standard curve – check your standard curve to make sure ‘BackCalcConc’ is the same as your input ‘Conc’, and that your curve looks correct.
- One way to calculate your concentrations is to take the ‘Value’ column, which is absorbance, plot this for the slope, and then back calculate the absorbance of your samples.
- Export your data
- Save your file. This will be in .sda format and will allow you to return to this specific run in the future to inspect run parameters
- To export the data, click on the 96-well plate in the top left corner of the screen and “export.” In the next window, click the experiment you are interested in exporting data for (usually this is just “Exp 1”), and it will save as a .txt format in the same name as your .sda file.
Calculations
- Preform a linear regression of the “Values” column versus the known concentrations of RNA in your standards.
- Subtract the “blank” value from your unknown RNA values
- Calculate the RNA concentration from the linear regression equation using your unknown values subtracted from the blank.
- First, account for the 1:100 dilution you performed to prep your plate by multiplying by 100.
- Second, change units (from ng/ml to ng/ μL) by dividing by 1000
Refractometer Modification
This modification comes from the Buckley lab
Items needed:
- Small sharp scissors
- AR200 Digital handheld refractometer (Cat. No. 13950000)
- Electrical tape
- 100 % isopropanol
- DEPC water
- Kim Wipes
- Cut out a circle of electrical tape with a smaller hole inside of it and place inside refractometer and over the edges (look at photo).
- If the black electrical tape is too big for the refractometer circle, you will get an error saying that the image is fuzzy. Cut down to appropriate size.
- Make sure the setting is nD-TC
- Note: this is very important. This measures the density (nD) and compensates for the temperature (TC) so that you have density readings at exactly 20C. You will do your spin at 20C, so you will need to know what the density of your gradients are at the same temperature you spin it at. The lab temperature is not static, so if you don’t have this setting on, your density values will be off by a decent margin and your SIP spins will be incomparable between different days.
- Clean the hole with 10 μL of 100% isopropanol, kim-wipe off the isopropanol. Add 10 μL of DEPC water, kim-wipe off the DEPC water. Add 10 μL of DEPC water.
- Hit calibrate, wait for the instrument to calibrate, then hit read. It should read exactly 1.3330. If it does not, kim-wipe the water and repeat step 4 again. It is important that the refractometer is calibrated consistently between SIP runs.
- If you want to remove the electrical tape, then just pull it off and clean the refractometer with 100% isopropanol and MilliQ water.
Gradient Salt Solution
Recipe from Dunford and Neufeld, 2010
Combine
- 50 ml of 1 M Tris-HCl
- 1 ml of 0.5 M EDTA
- to 400 ml of water.
- Dissolve 3.75 g KCl, then add ddH2O to 500 ml.
- Filter-sterilize and autoclave.
The final solution is 0.1 M Tris, 0.1 M KCl and 1 mM EDTA.
Gradient Setup
Protocol is a derivation of the Huber lab protocol, Lueders 2010, Buckley Lab RNA SIP Protocol, and Roey Angel’s protocol
Day Before Gradient Setup
The day before you setup a spin, hydrate your 50 g powdered CsTFA with 13 mL of GSS. Shake vigorously and allow to come to room temperature for several hours to overnight (I do overnight). Measure the RI of the GSS you hydrate the CsTFA with (after the calibration). GSS can increase in density due to evaporation over time.
- Note: CsTFA is extremely hydroscopic (i.e. loves water) and will become crusty if you leave the powder out after opening the bottle (i.e. absorbs water through humidity). This step ensures consistency between SIP spins over time.
Day of Gradient Setup
- Rinse Beckman tubes with 500 μL of 70 % molecular grade ethanol. Hang upside to dry and tap every so often to get out the excess ethanol.
- Label 15 ml tubes with sample names. Do not run more than 6 - 8 samples. You may want to include one No RNA control at first.
- Set your refractometer to nD-TC and do the isopropanol/water/water/1.3330 calibration described above. Do not move forward until the refractometer is calibrated to 1.3330.
- nD-TC allows you to read the refractive index at 20C, which is what you run the ultracentrifuge at. Densities will change with temperature, so the density at room temp (~23C) will be different than 20C. Calibrate the refractometer with 75 μL of DEPC water. It should read 1.3330 +/- 0.0002. If it is off, clean the refractometer with 100 μL of 100% isopropanol and read it again.
- For 6 samples - Add 30 mL GSS+CsTFA to a 50 mL tube. Measure the RI and bring the solution to 1.3765 with GSS (usually about 3 mL).
- Filter sterilize 5.1 mL into 15 mL tubes and add 185 μL of formamide. Record the RI for each of the tubes after adding formamide (important for Lueders 2010 correction factor).
- Add up to 1 mL minus volume of 750 ng of RNA of the Gradient Salt Solution. E.g. If you are going to add 20 μL of RNA, then add 980 μL of Gradient Salt Solution. Recipe for solution can be found in Neufeld et al. 2010, JoVe article.
- I usually start with 500 μL GSS minus the volume of RNA and then adjust the density with more GSS (too high) or CsTFA solution (too low) in 50 μL increments.
- Add 750 ng of RNA. Can add 500 - 750 ng, but try not to go lower.
- Mix gently by inverting. DO NOT vortex.
- Using 10 μL from each tube, measure Refractive Index (RI) on the refractometer. Each tube should measure 1.3729±0.0002. This may be adjusted to optimize peak fractions if needed.
- If RI is too high or too low, adjust using CsTFA buffer (if too low) or Gradient Salt solution (too high). Add 50 - 100 μL to adjust, then mix and re-measure. DO NOT proceed until each tube is within range.
- Also, measure pure CsTFA each time to check if there are changes in density (not usually needed).
- Fill each 5.1 mL Beckman centrifuge tube using a pipette. Fill to the bottom of the neck. Do not overfill! Place black caps on top. You should notice the bottom half of the black cap is wet. It is too full if fluid is higher or is coming out of the tube. Level is too low if the fluid does not touch the cap.
- Determine the mass of each filled tube on the scale by placing each tube in the brown flask stopper with the hole in it and measuring its weight. Make sure the scale’s bubble is in the center before weighing. Each tube should have a partner with comparable mass (within 0.01 g). If one tube is greater than 0.01 g heavier than the other, add GSS in 10 μL increments until the masses are the same.
- After looking up the proper configuration for the number of tubes you have (below), pair up tubes with similar masses and load into the rotor. MAKE SURE IT IS BALANCED. Place weighted gold caps on top of tubes (skinny side up; press all the way down), then thread in the caps, white o-ring down. Use a ratchet/torque wrench to tighten to ~ 40 lb.
- Place the rotor in the ultracentrifuge in Harriet’s Lab. Make sure to only set the rotor down in the base to avoid scratching the code on the bottom. Run @37,000 rpm, 20 °C, for 65 hours. Make sure the program has NO BRAKE (setting 10) for deceleration option.
- After setting up your spin, calculate the total volume of the tube (i.e. 5.1 mL + .850 mL) and the amount of formamide use. Multiply those volumes by 0.1 and make a GSS + formamide solution in a 1.5 mL tube from those calculated values. Measure the RI (after calibrating) and subtract that value from 1.3333 for the Buckley lab correction factor.
Note: when the cycle is finished, it will not say “end” or “complete;” it will look like you didn’t start it at all. This is why it is important to check on the ultracentrifuge every day that it is running, to make sure it is still running correctly.
Fractionation
Items needed
Protocol
- Prepare 50 mL tubes of a) 70 % ethanol and b) 100% isopropanol, then place in the -20 freezer for later.
- Make ~50 mL of colored MilliQ water with Resazurin dye (add 100 μL).
- Rinse low-flow tubing with MilliQ water and 70% EtOH before use.
- Set up a ring stand with a test-tube clamp to hold the Beckman tubes.
- Sterilize deep well plates and sealing mat before use. Label both the plate and sealing mat with cryolabels.
- Take the rotor out of the ultracentrifuge, place it in the base, bring it to your bench, unload tubes, and place them in the black tube rack.
- Set the pump to 0.45 mL/minute.
- Prime tubing with colored MilliQ water.
- Screw needle on end of tubing and prime again.
- Place a 5 mL Beckman tube gently in the clamp (do not squeeze the tube) and ready the plates. Make sure A1 is the FIRST well filled - this is the heaviest fraction.
- Puncture the bottom of the 5 mL tube with a needle. Place a piece of tape over the hole, so that none of the liquid spills out when you make the next hole in Step 13.
- Place the plate under the tube rack.
- Puncture the top of the 5 mL tube with the primed needle that is now attached to the tubing.
- Set a stop watch on your phone. Simultaneously press start on the timer and the pump.
- Shift rack every 30s to collect a new fraction. Record where each fraction is on the sheet. Look at the template below for an example of how to set up a plate.
- After 11 minutes, you have collected all 22 fractions. Turn off the pump.
- Now cut a piece of sterile sealing film (like BioRad Sealing Film B) into thirds. Cover the fractions you just collected with the third piece of sealing film.
- Collect the second set of fractions on the other side of the place (i.e. RNA2 from the template) starting with well H12.
- Once you have finished one plate, remove the sealing film and cover with the sealing mat, careful to match up the A1 on the sealing mat and the plate.
- Measure and record the RI for each fraction.
- Make sure the setting is at nD-TC and repeat the isopropanol/water/water/calibrate/1.3330 every time before you start reading a set of fractions. Do not proceed if the DEPC water is not at 1.3330. Start with the lightest fraction first (i.e. 22) to prevent residual denser fractions from interfering with your readings.
- Add 10 μL of DEPC water or sample. The middle fractions (~11-13) should be ~1.3729, or approximately the same density as the starting solution. If this is not the case, the samples may need to be re-run.
- Repeat fractionation for each sample. Repeat the isopropanol/water/water/calibrate/1.3330 step before each set of fractions.
- Add 1X volume ice cold 100% isopropanol to each well. Invert gently 30 times and place in the -20 °C overnight.
- Clean tubing by flushing with MilliQ and then 70% Ethanol. Let the ethanol drain out and dry by hanging the tubing from the center over a paper towel.
RNA Precipitation
Before precipitating – if you have an odd number of plates – make a “buddy” plate for the isopropanol and ethanol washes using water. Make sure that it weighs about the same as the plate with CsTFA. Note: CsTFA is heavier than water, so you may need to add like 1.1-1.3X more water to each well to compensate for differences in density.
- Spin the plates in the centrifuge at 4 °C, 4250 rpm for 45 minutes.
- Take off the lids (remember to label the lids) and very quickly dump the isopropanol wash into the trash can. While the plate is still upside down, place a paper towel over it to remove the residual isopropanol.
- Flip the plate back over. Add 150 μL of ice cold 70% ethanol to each well.
- Spin at 4 °C, 4250 rpm for 25 minutes.
- Take off the labelled lids and very quickly dump the ethanol into the trash can. While the plate is still upside down, place a paper towel over the wells to remove the residual ethanol.
- Dry the RNA by letting the plate sit out for ~30 minutes.
- Hydrate the RNA by adding 20 μL of 1X TE. Vortex and spin down at 4250 rpm for one minute.
- A freeze-thaw always works best for me before ribogreening. Let the plates sit in the -80 °C overnight before ribogreening.
Quantification/Correction Factors
Ribogreen all fractions to determine concentrations. Using the protocol described on p. 4-5 but with 2 μL template and modified plate setup to allow easier quanitifcation from 96 well plates.
We use correction factors to compensate for differences in the GSS and formamide between runs. This was not done previously with the liquid CsTFA solution. There are two ways to correct currently. The first is from Lueder’s, which is not a published protocol. The second is from the Buckley lab. At this point and time (Sept 2022) there has been no recent research published on correction factors for RNA SIP (except Buckley’s).
Leuders
Use CsTFA standard curve from figure 2B (below) Lueders 2010 to convert RIs into buoyant density and apply the correction factor.
Correction factor:
(density of CsTFA + Gradient Buffer + formamide: 1.98) - density of CsTFA solution in water adjusted to RI of bought solution.
Therefore, if the RI of your CsTFA + Gradient Buffer + formamide was 1.3783, then the buoyant density of is 1.91549008. Your correction factor would be 0.06450992 (1.98-1.91549008). The formula in excel is: 1893.982POWER(RI_CsTFA+GSS+Form,2)+5230.8018( RI_CsTFA+GSS+Form)-3609.6804).
Buckley
This is derived from the Buckley Lab SIP Protocol
We calculated the correction factor in step 13 of Gradient Setup. Subtract that value (i.e 0.0051) from all of your RI values from your SIP spin and then calculate using the equation in the link:
RI corrected = RI observed – (RI Buffer - 1.3333)
RI observed = Either the GSS + CsTFA or fractions
RI Buffer = GSS + Formamide (no CSTFA)
Buoyant Density = 163.559 – 262.271*(RI corrected) + 105.281*(RI corrected)^2
KAPA SYBR Fast One-Step RT-qPCR